DataMan 280 Installing C-Mount Lens
- Unscrew the four M2x12mm Phillips Pan head screws, then take off the metal cover from the reader.
- Remove the illumination module.
-
Unscrew the two M2x5mm Phillips head screws, remove the 6.2 mm lens mount and the 6.2 mm lens.
-
Place the C-Mount cover on the front of the device, and tighten the screws.
Note:Observing the tightening sequence below, tighten all six M2 x 12 mm Phillips Pan screws to 12 Ncm using a torque wrench.
- Place the lens cover adapter plate on the C-Mount cover, and tighten the screws.
- Thread the lens into the reader.
- Screw the lens cover onto the lens cover adapter.