Skip to Main Content

Changing DataMan Verifier Light Settings via DMCC

DMCC documentation is lagging details on how to apply changes in ISO29158 light settings for individual Read Setups.

17/02/2026

Details

Controlling light settings per read setup using DMCC commands

Overview

In more advanced applications, customers may want to dynamically switch lighting configurations across multiple Read Setups using a PLC or PC via DMCC commands. Although the command ||>SETUP.ENABLE-SINGLE selects a specific Read Setup for decoding, it does not redirect subsequent light-configuration commands to that setup.

This article explains the correct procedure for addressing light settings within a specific Read Setup.

 

Issue description

A common misconception is that sending:

||>SETUP.ENABLE-SINGLE <n>
automatically causes:

||>SET LIGHT.DIRECT ...
to apply to the same Read Setup.

This is not the case.

||>LIGHT.DIRECT will still target Read Setup 0 unless an additional command is issued.

 

Required command: SETUP.PROG-TARGET

To modify lighting parameters inside the Read Setup you intend to program, you must also set:

||>SET SETUP.PROG-TARGET <n>
This command directs configuration changes—such as lighting—to the specified Read Setup.

 

Correct command sequences

Case 1: Modify light settings in read setup 0

||>SETUP.ENABLE-SINGLE 0
||>SET SETUP.PROG-TARGET 0
||>SET LIGHT.DIRECT OFF OFF OFF OFF ON ON ON ON OFF
||>TRIGGER ON

Case 2: modify light settings in read setup 1

||>SETUP.ENABLE-SINGLE 1
||>SET SETUP.PROG-TARGET 1
||>SET LIGHT.DIRECT OFF OFF OFF OFF OFF OFF OFF OFF ON
||>TRIGGER ON

 

Summary

SETUP.ENABLE-SINGLE controls which Read Setup is used for decoding.

LIGHT.DIRECT always defaults to Read Setup 0 unless redirected.

Use SETUP.PROG-TARGET <n> to write lighting or configuration changes to a specific Read Setup.

Both commands must be used together for correct behavior.