DMST versions post 23.x.y will not detect the reader on Ethernet anymore. Older version did not have this problem.

18/02/2026

Details

In a recent use case the DHCP server on the network was configured in the IP range 169.254.X.Y and the 24.X.Y and 25.X.Y releases of DataMan Setup Tool did not auto-detect the reader on the Home Tab anymore.

Background:

The 169.254.0.0/16 IP range is reserved by RFC 3927 for Link-Local addressing. Devices use this range automatically when no DHCP server is available. This mechanism is commonly known as APIPA (Automatic Private IP Addressing) in Windows, or more generally "Link-Local Addressing."

Issues when using this range via a DHCP server:

  1. Misuse of a reserved range:
    If a router or DHCP server actively assigns IPs from this range, it violates the RFC standard and contradicts the expectations of many devices and networking protocols.
  2. Problems with UDP-based discovery protocols:

    Many devices (e.g., printers, cameras, IoT devices) ignore or mishandle the 169.254.x.y range for regular communication, as it is typically treated as a fallback only. As a result, UDP broadcasts (e.g., SSDP, mDNS, Bonjour, Profinet, etc.) often don’t work reliably in this IP range.

  3. Routing and gateway issues:
    The Link-Local range is non-routable by design. If you use it in your network, communication across subnets or via gateways will not work, as routers do not forward Link-Local traffic.
  4. Troubleshooting becomes confusing:
    If 169.254.x.y addresses are assigned intentionally, it’s no longer clear whether a device is using this address due to a network failure (e.g., no DHCP) or because it was explicitly configured to do so. This makes diagnostics more difficult.

Conclusion:

Yes, using the 169.254.0.0/16 range via DHCP is problematic. It can cause serious issues with UDP discovery protocols, routing, device communication, and troubleshooting.

Recommendation:

Use one of the RFC 1918 private IP address ranges instead:

  • 192.168.0.0/16
  • 10.0.0.0/8
  • 172.16.0.0 – 172.31.255.255

These ranges are specifically designed for private networks and are widely supported and expected by networking equipment.